ProChem SSW (Steam Sterilization Integrator)
- Class 5 Integrator
- Moving front (wicking) Integrator
- Eliminates need for interpretation
- Catalog # CI-SSW
- 250 Integrators/bag
- Expires 5 years from date of manufacture
ProChem SSW Steam Sterilization Integrator is a single use Class 5 Integrating Indicator which complies with ANSI/AAMI/ISO 11140-1:2005. It integrates three essential criteria for proper steam sterilization: time, temperature, and steam. When processed along with a load it indicates, with a margin of safety, whether the proper steam sterilization criteria have been achieved.
ProChem SSI (Steam Class 4)
- Class 4
- Measures time, temperature and presence of steam
- Catalog # CI-SINT
- 250 indicators/box
- Expires 3 years from date of manufacture
- White to Black color change is not reversible
- Instant visual verification of adequate steam penetration into the pack/pouch
- Easy to Interpret
- 3 minutes at 270-275°F (132-135°C)
- An integral part of a comprehensive sterility assurance program
- Non-reversible color change
- International use only
Use ProChem SSIs to verify adequate steam penetration into every pack/pouch. The indicator ink changes from white to black during steam sterilization. Tests have shown that ProChem SSIs complete their color change at a time equivalent to or beyond the time needed to kill Biological Indicators. (Data available upon customer request) Use according to instructions. ProChem SSIs are packaged 250 per box, with 16 boxes per case.
This product is manufactured in an FDA registered facility pursuant to FDA Quality System Regulations (QSRs). Use ProChem SSIs in conjunction with BIs as described in current AAMI Standards and Recommended Practices: Good Hospital Practice: Steam Sterilization and Sterility Assurance. Use according to instructions on box.
O.K. Cycle
- Catalog # CI-OK
- 250 indicators/bag
- Expires 2 years from date of manufacture
- International use only
The Mesa O.K. Cycle multispot Chemical Integrator strips are designed for use in pre-vacuum steam sterilizers operating at 121° C and at 134°C (ISO 11140-1:2005 compliant as a Class 5 Integrator). Used as directed the Mesa O.K. Cycle integrator strips give visual confirmation that sterilization conditions were met at point of placement in the sterilizer chamber. During steam sterilization, the four indicators spots change from purple to green.
For 121°C cycles, verify that indicator spots 1, 2, and 3 have changed from purple to green. Spot 4 will change to green in sterilization cycles exceeding 35 minutes.
For 134° C cycles, verify that indicator spots 1 and 2 have changed from purple to green. Spot 3 changes to green after spot 2 giving the user more information about the cycle. See critical parameters chart. Spot 4 will change to green in sterilization cycles exceeding 12 minutes.
O.K. Cycle Technical Specifications
Spot
1
2
3
4
121 Degrees C
8 minutes
11 minutes
17 minutes
30 minutes
134 Degrees C
3.25 minutes
4.5 minutes
7 minutes
11 minutes